Previous Page  15 / 16 Next Page
Basic version Information
Show Menu
Previous Page 15 / 16 Next Page
Page Background

Ready for the future,

caring for the customer

FPA-

128

M TECHNICAL SPECIFICATION

GENERAL

Electronics

Compatible with SONIA architecture

UT channels

32x128 multiplexed channels (TX/RX separated)

UT techniques

Pulse-echo, through-transmission, tandem

PULSER

Type

Negative square wave

Voltage

-15 to -100 V, 1V steps.

Width

20 to 500 ns, 2.0 ns steps

Maximum PRF

20 KHz

Fall time

<6 ns

Impedance

<15 Ω

RECEIVER

Input range

0.8 Vpp

Bandwidth

0.45 to 28MHz @-3dB

Gain range

0 to 80dB, 0.1 dB steps (analog gain)

0 to 40dB, 0.1 dB steps (digital gain)

TGC

60 dB maximum range, 16 points, 16 ms extent,

32 ns resolution, 40dB/us slope.

BEAM FORMER

Focal laws

Up to 4096, 128 scans.

Electronic Scan types

Linear, sectorial, complex

Focusing

Dynamic Depth Focusing (DDF)

DIGITAL FUNCTIONS

A/D sampling rate

125 MSPS, 12 bits (internal processing @18 bits)

Sampling decimation factor: 1 to 16

Gates

4 gates, max + 8 first echoes per gate.

1 gate for interface echo synchronization.

Signal modes

RF, True Envelope, Rectified (full, +/-),

Logarithmic (digital)

Frequency Filters

Digital programmable IIR type, low-pass,

high-pass, band-pass.

Noise reduction filters

Averaging

Anti-impulsive noise

Post-Rectsmoothing filter

0 to 100% smoothing level control

Other digital functions

Real time alarms associated to echoes in gates.

Signal inversion.

Signal compression factor up to 64.

OTHERS

UT connectors

I-PEX Minidock

General I/O

4 digital inputs, 2 digital output (up to 24 V)

optocoupled.

Fiberoptic port

LC-Duplex optical connector (1 Gbitfull-duplex)

Dimensions

166x 111x 64 mm (DxWxH).

Power

24 Vdc, 25W typ.

Accessories

IPEX to Hypertronics connectors converter

IPEX splitter (2x64 channels) for dual probe

applications